Chemicals List for Analytical Performance (CLAP)

Catalog of Regulatory Science Tools to Help Assess New Medical Devices 

 

The CLAP tool is chemical dataset intended to help analytical laboratories demonstrate their ability to detect the broad range of chemicals that may be encountered in a non-targeted analysis as part of a comprehensive extractables leachables program.

 

Technical Description

CLAP is a collection of over 100 example chemicals that have been characterized in CDRH laboratories and are provided with the following information:

  • Trivial/ common name or acronym
  • Full name and CAS#
  • Empirical formulae
  • Double bond equivalent (DBE), range 0-25
  • Molecular Weight (g mol-1), range 102 to 1178
  • Boiling point (°C), range 80 to 779
  • pKa, range -9.1 to 18.25
  • LogP, range -0.2 to 20.8
  • Relative Response Factor (RRF) for GCMS, determined against the internal standard (code A5) 2,4-Di-t-butyl phenol at a concentration of 5µg ml-1 for all compounds.

 

 In addition, the specific protocols used to determine the RRF for each chemical are provided. 

Intended Purpose 

The information is intended to be used by analytical chemistry labs to determine their ability to detect chemicals of sufficiently broad physicochemical properties.  The chemicals have been specifically chosen to represent broad physicochemical properties and are readily available.

Using the GC/MS or LC/MS protocols provided or variations that have already been developed by individual laboratories, detectability of the chemicals is assessed by the chromatographic relative response factor. Demonstration that the techniques used can detect chemicals across a broad set of all physicochemical properties of chromatographic importance is an important surrogate measure for the completeness of non-targeted analysis during chemicals characterization of medical device extractables. 

The set of RRFs measured can then be used to populate a coverage map comparing the RRFs obtained across the breadth of physicochemical properties. Such data may be useful in chemical characterization as described in the International consensus standard ISO10993-18 “Biological Evaluation of Medical Devices. Part 18: Chemical characterization of medical device materials within a risk management process.”

As a part of premarket biocompatibility evaluation for medical devices, appropriate chemical characterization is an essential step to inform the toxicological risk assessment (TRA) process and determine the benefit risk profile for a medical device.

Use of the coverage map approach outlined in this tool increases the confidence that all relevant chemicals have been detected, allowing for identification and TRA to be conducted.

Testing

  • All testing was undertaken in the OSEL labs using an Agilent Technologies 6890N gas chromatograph coupled with an Agilent Technologies 5975B mass spectrometer.
  • The columns used were an Agilent HP-5MS (30m X 0.25mm i.d., 0.25µm) and an Agilent DB-5HT ((30m X 0.25mm i.d., 0.1µm). Helium (99.999% purity) was used as the carrier gas at a flow rate of 1mL min-1.
  • Injection volumes were 1µL or 2µL.  Other parameters were:
ParameterValue
Transfer line temperature300 or 325 oC
Initial temperature40 oC
Final temperature310 or 340 oC
Ramping rate7.5 or 15 oC min-1
Mass range35 to 650 m/z
Solvent delay2.5 min

Limitations

  • Use of the chemical data set does not guarantee that an appropriate range of detectability has been necessarily achieved.
  • RRFs are highly dependent on the specific protocols used for GCMS and LCMS and should always be specified when using the chemical dataset to generate a coverage map. 
  • The chemical data set has been designed to cover a specific range for each physicochemical property. The specific ranges should not be interpreted as a limit on the breadth of physicochemical properties of extractables that may be relevant for TRA.